Re: [AMMRL] Bruker BBO diff probe with water cooler problem

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 14 Aug 2024 23:07:44 +0200

Hi,

did you check whether the Great60 PFG amp maybe permanently outputs some DC
current? You should see this by observing a high-resolution spectrum (say,
lineshape sample or the like) with the PFG cable connected versus disconnected
(and amplifier online) - if the lineshape gets bad by connecting the PFG amp
then it may have a permanent DC offset and may require fixing.

I guess you checked for proper water flow going through the probe, yes? (maybe
introduce a flow meter or plug the "in" port of the water chiller and have the
tube coming from the probe pour the water directly back into the reservoir for
you to see)

Best regards
Bert

> Am 14.08.2024 um 20:19 schrieb Wang, Tianzhi <tiawang_at_utmb.edu>:
>
> Dear All
>
> I have a Bruker BBO diff probe with water cooler. But it never worked with
> a water cooler. It worked fine with the Great 10 gradient amplifier. The
> problem happened when I switched to Great 60A with a water cooler system.
> The amplifier was stopped by the error " the temperature is too high",
> actually, the temperature jumped quickly between 26-49C every 5 mins as I
> observed, since the gradient amplifier was stopped by the temperature 50 C
> or above. It was leaking water when Bruker installed it twice and then sent
> it back to Bruker for the repairing three times for almost one year. I can
> only use it maximum 3 days after each repairing. The problem happened at first
> by the error "the temperature is too high" and later developed to the message
> "output power fault". Then the gradient amplifier stopped completely even
> with 1% gradient. From my understating, "output power fault" should be
> the output voltage is too high.
> It would be great if you could share your BBO diff probe with me.
>
> Best,
> Tianzhi
>
> -------------------------------------------------------------------
> Tianzhi Wang, Ph. D.
> NMR Manager
> Sealy center for Structural Biology & Molecular Biophysics
> University of Texas Medical Branch,
> 301 University Bvld, NMR dockside building, Rom 1.104
> Galveston, TX 77555-1157
> Email: tiawang_at_utmb.edu <mailto:tiawang_at_utmb.edu>
> (409)7476821 (O)
>
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1498): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1498__;!!PvDODwlR4mBZyAb0!RTFXjoJe7I_8NQbrYY99rhIGMqrf0E=
9M_HuoFwvObswi1xVK2RBI8UfMNFszvZQt0IiKiF0nWEUj8EO2qD7jjg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107900562=
/7559972__;!!PvDODwlR4mBZyAb0!RTFXjoJe7I_8NQbrYY99rhIGMqrf0E9M_HuoFwvObswi1=
xVK2RBI8UfMNFszvZQt0IiKiF0nWEUj8ENFutJMhQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Aug 14 2024 - 14:14:10 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST