Re: [AMMRL] Off axis shims bad after 24 hours on 40-year-old Oxford magnet #Bruker #Hardware

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Wed, 14 Aug 2024 12:25:30 -0400

Hello Wendy,

Years ago when we had a magnet that we suspected might quench, I was
advised to remove not only the room temperature probe, but also the shim
stack.  If/when the magnet quenches, the bore of the magnet will get
extremely cold, leading to considerable ice/condensation buildup. 
This is especially true when the quench leads to a loss of vacuum in the
magnet can.

Whether or not the bottom flange blows out when the magnet quenches, I
would not want to subject the probe to the extreme temperature drop and
resultant condensation.  So I would definitely warm up and remove the
cryoprobe.

If you would still like to make periodic measurements, then you could do
so with a room temperature probe.

If you also remove the shim stack, then you won't be able to take any
NMR measurements.  It's up to you to determine if being able to take
measurements is worth the risk to the shim stack electronics.
NOTE:  If you do decide to remove the shim stack, be sure to power
down the shim controller - a.k.a. the BSMS rack - before doing so.

For me, I would be inclined to leave the shim stack in, and load an RT
probe.  That way you can continue to monitor how the magnet is doing.

Good luck,
Bob.

On 2024-08-14 11:32 a.m., Wendy Breyer via groups.io wrote:
>
>
> Thanks, everyone, for the suggestions.  I'm currently working through
> them but I have some questions for those of you who may have had an
> old Oxford magnet quench:
> Our magnet was retrofitted with special bolts at the magnet flange. 
> I am not 100% certain but I believe this was done to allow the bottom
> plate of the magnet to drop in the case of over-pressurization as
> would be expected in a quench.  I'm a bit concerned for our newish
> cryoprobe that is currently installed... if the bottom plate drops,
> how forceful is it? Will it damage the cryoprobe?  Should I remove the
> cryoprobe ASAP because of the possibility of damage?  Unfortunately I
> have no idea how likely a quench is and how forceful the dropped plate
> will be...  Any advice or experience would be welcome!
> Thanks again,
> Wendy
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
--------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With OneSpoon"
wampum agreement.



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1495): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1495__;!!PvDODwlR4mBZyAb0!WAoJByrfVdQ4WZLLVf_mrkoaxcOIHF=
GZUr1DhLWANPuTdatC7rzjHCyv-c-2gBMt2N568sxGa0pL7jkU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107818605=
/7559972__;!!PvDODwlR4mBZyAb0!WAoJByrfVdQ4WZLLVf_mrkoaxcOIHFGZUr1DhLWANPuTd=
atC7rzjHCyv-c-2gBMt2N568sxGa63ug0F-$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!WAoJByrfVdQ4WZLLVf_mrkoaxcOIHFGZUr1DhL=
WANPuTdatC7rzjHCyv-c-2gBMt2N568sxGayT8xiAZ$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!WAoJByrfVdQ4WZLLVf_mrkoaxcOIHFGZUr=
1DhLWANPuTdatC7rzjHCyv-c-2gBMt2N568sxGa1CBbHPS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 14 2024 - 09:25:42 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST