Re: [AMMRL] Question about LHe cryoprobe costs/He recovery

From: Clemens Anklin via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Thu, 31 Aug 2023 11:05:52 +0000

Dear Jessica,

Helium cryoprobes do not use liquid helium at all. They work with a closed
loop helium compressor and cooling system. You will need high purity ( 6 nines)
Helium gas for installation and operation. Consumption depends on how often
you remove and reinstall the probe. The recurring cost with a cryoprobe is
the periodic maintenance of the coldhead.

With best regards

Clemens


On 8/30/2023 8:48 PM, Roberts, Jessica wrote:
**EXTERNAL EMAIL**
Hi everyone,

I am in the process of determining the most appropriate cryoprobe to purchase
and one of the major driving forces is that of the operation/upkeep costs. I
have a handful of biochemical users, as well as other advanced organic users,
that would greatly benefit from a LHe cryoprobe and personally that is my preference.
However, I was curious to see a few things from others that would help navigate
this decision:


  1. What are the yearly costs related to the upkeep of the LHe cryoprobe?
If the cost can be broken down to compare between general upkeep and He costs,
that would be super helpful!
  2. I know that is going to depend on usage and operation temperatures, etc.,
but how much LHe do you feel your probe goes through yearly?
  3. Is there a way to recover the He boil off from the probe? I am in the
process of installing a He recovery hose to the NMR He stack already, not
sure how that would work to recover from probe (I apologize if this is a
silly question).

Thanks!!!
Jessica

Dr. Jessica M. Roberts, Ph.D.
Assistant Professor of Instruction
NMR Facility Manager
Department of Chemistry and Biochemistry
Ohio University
245 Chemistry
Athens, OH 45701
740.593.0038
jroberts_at_ohio.edu<mailto:jroberts_at_ohio.edu>
--




--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#491): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/491__;!!PvDODwlR4mBZyAb0!XePvpWC9AHaxv0OEiO0BrFW4_bR6Jw55=
ARFdGAi4JH6613a_vERDRbzNvQH-26N9blnzYON-opiEMlLG8JRlWtvIuPWB$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101067096=
/7559972__;!!PvDODwlR4mBZyAb0!XePvpWC9AHaxv0OEiO0BrFW4_bR6Jw55ARFdGAi4JH661=
3a_vERDRbzNvQH-26N9blnzYON-opiEMlLG8JRlWoxzrBuO$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Aug 31 2023 - 04:07:03 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:56 MST