Hi Ronald
Over the years I have had to re-vacuum transfer lines but never had one repaired
or purchased one. Usually you can find an adaptor that will allow you to hook up
to a KF vacuum connection. Older system usually had this in the magnet spare parts
box. We use a turbo pump assembly from our EPR pull new vacuum on lines and cryogen
dewars. Can send a picture of what one looks like if you want.
Greg
-=-=-=-=-=-=-=-=-=-=-=-