Hello Spinlanders

Thank you to everyone for your quick responses!

Over whelming majority of you are just using natural gas meters and the most popular seems to be the AL-425 by American Meter. This is the one we were looking at for monitoring gas from fills but the specs indicate it will not measure regular boil off flow levels. But several of you indicated that this meter does just fine with monitoring all the gas flows.

SO thank you very much for your replies we will be purchasing several of these meters.

Greg

----------------