David Rice's answer is better than mine, as it has the same effect of
eliminating the early part of the fid but in real time instead of post-hoc.
However, you have to fine-tune that gap between the detection pulse and the
acquisition so as not to throw away too much good signal.
Bill
On Aug 18, 2015 1:23 PM, "Kirk, Megan (MM)" <megan.kirk_at_sasol.com> wrote:
>
> Hello everyone,
>
>
>
>
>
> How does one prevent a rolling baseline in solid state 31P NMR on a narrow
> bore Varian 600 MHz, Unity Inova? I know I can remove it afterwards by
> integration and then using dc and bc. How do I prevent it in the first
> place?
>
>
>
> Thanks for your help,
>
>
>
>
>
> Megan Kirk
>
>
>
>
>
>
>
> [image: sasol_logo]
>
> 1 Klasie Havenga, Sasolburg, 1947
> PO Box 1, Sasolburg, 1947
> www.sasol.com
>
>
> *Megan Kirk *Scientist
>
> R&T Operations & Analytics
>
> Tel
>
> +27 16 960 2025
>
> Fax
>
> +27 11 522 3906
>
> Mobile
>
> +27 82 5743428
>
> E-mail
>
>
>
> megan.kirk_at_sasol.com
>
>
>
>
>
>
>
>
>
> ------------------------------
> NOTICE: Please note that this eMail, and the contents thereof, is subject
> to the standard Sasol eMail legal notice which may be found at:
> http://www.sasol.com/legalnotices
> If you cannot access the legal notice through the URL attached and you
> wish to receive a copy thereof please send an eMail to
> legalnotice_at_sasol.com
> ------------------------------
> disccsa41a3
>
Received on Wed Aug 19 2015 - 09:05:22 MST