AMMRL: Answer to "Can TopSpin 1.3 run on Centos 5?"

From: Wycoff, Wei G. <WycoffW_at_missouri.edu>
Date: Thu, 31 Jul 2014 22:05:59 +0000

Dear AMMRL members,

Thanks to many of you who answered my question concerning whether TopSpin 1.3/CentOs 5 combination would work well on our DRX300 NMR. I am happy to learn from your responses that this combination has worked well for all of you. A few things don't work initially (some AU programs won't compile and plot-editor won't start) but the solutions can be found on Bruker Knowledgebase website under item #9585, #8702, #9577 and #9641. TopSpin 1.3.pl10 now is working with CentOs 5 on our 17-year old DRX300 spectrometer. (Also, the two SCB boards and the LCB board in our console had to be upgraded to a higher ECL level in order to run TopSpin 1.3.pl10.)

Thanks for your help again!

Wei

Wei Wycoff
University of Missouri
NMR Facility
125 Chemistry Building
601 S. College Ave.
Columbia, MO 65211

Office: 573-884-7807
Lab: 573-882-3291
Fax: 573-882-2754
Email: wycoffw_at_missouri.edu


-----Original Message-----

> Dear AMMRL members,
>
> Have any of you installed TopSpin 1.3 on a PC/Linux Centos 5? If you
> have, is this combination working well? We would like to install this
> combination on our DRX300 spectrometer.
>
> Thanks,
>
> Wei
Received on Thu Jul 31 2014 - 12:06:09 MST