I'm just posting this as an FYI in case anybody else encounters this
problem.

Recently, we shutdown a Bruker cryoplatform for a planned power outage in
our building. Upon trying to restart, the power button would not engage. I
would push it, the lights would turn on, but upon releasing the button,
power would be cut again. The contact block for the power button had
failed. It was a fairly cheap/simple repair, though a bit tricky to remove.
The part description says "lifecycle of 1 million switching cycles" so it's
possible that I am the only one who will ever encounter this problem, but
in case somebody else does, here is the part info and a link to it at

Radwell:

Schlegel BFL5 Illuminated Contact Block
https://www.radwell.com/Buy/SCHLEGEL/SCHLEGEL/BFL5

PXL_20260409_140557343.MP(1).jpg]


_____________________________________
Todd Rappe
Minnesota NMR Center <nmr.umn.edu>
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3374): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3374__;!!PvDODwlR4mBZyAb0!QuRqpS_u-Hm75o1JthfKYVyLmjgP-7=
xRzbuRbCkP1HJHL-HPjniMDPH8hRyN89H-6R3NrhwzioCntX52IqekgmX-xSEG$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118743388=
/7559972__;!!PvDODwlR4mBZyAb0!QuRqpS_u-Hm75o1JthfKYVyLmjgP-7xRzbuRbCkP1HJHL=
-HPjniMDPH8hRyN89H-6R3NrhwzioCntX52Iqekgsbg8W1L$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






PXL_20260409_141051286.NIGHT.jpg
(image/jpeg attachment: PXL_20260409_141051286.NIGHT.jpg)

PXL_20260409_140557343.MP_1_.jpg
(image/jpeg attachment: PXL_20260409_140557343.MP_1_.jpg)

Received on Thu Apr 09 2026 - 07:38:42 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST