Firmware updates for DRU and CCU went normally. I checked all of the boards
for the BSMS in service tool and none required downloading

However, now I am in a situation where BSMS service tool initiates itself any
time I do rsh, zg, etc. This loses my shim file and turns the sweep back
on (I am running solids).

So, I am not sure what needs updating/adjusting to get BSMS to stop resetting
itself. I did reboot the console and go through cf again.

Thanks,
Sarah


-=-=-=-=-=-=-=-=-=-=-=-