Re: [AMMRL] Bruker cryoprobe repair

From: Bob Berno via groups.io <bberno=mcmaster.ca_at_groups.io>
Date: Mon, 6 Apr 2026 13:43:41 -0400

Hello Jeff,

I suspect that the capacitor in your 1H tuning channel is shot. Because
this is a helium cryoprobe, you are somewhat limited in your options.
I fear that the probe will have to be sent away on an extended holiday.

In the meantime, you will have to suffer through using a traditional
room temperature probe (unless you happen to have a spare cryoprobe).

800MHz isn't exactly common, but you might be able to borrow (or rent) a
TCI cryoprobe from someone while you wait for your 19-year old probe to
be repaired.

There may be some third party options you can consider if you are unable
to send your probe back to Bruker for repair.  Where are you located?

Bob.

On 2026-04-06 11:35 a.m., Jeff Ellena via groups.io wrote:
> Hi All,
> Our 19 year old Bruker TCI 800S4 H-C/N-D cryoprobe has a tuning
> problem. The 1H channel can be tuned and matched only between 819 MHz
> and 821 MHz, BF1 is 800.280. All other channels tune and match
> normally. I removed to ATM unit and replaced it with a manual tuning
> unit and observed the same tuning problem. I removed the cryoprobe and
> replaced it with a TXI (room temperature) probe and 1H tuning and
> matching were normal. Bruker said the probe is unrepairable due to
> lack of parts availability and age.
> Does anyone know if there is a NMR probe expert who could repair the probe?
> Best,
> Jeff
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686

---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3356): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3356__;!!PvDODwlR4mBZyAb0!XpnKKmgtDbeWGvhHcYMtCjglwPRF-k=
jFEc9xwccJVSB_gqW6A_8gwdrAfS6U8pJgPlCMqgwYFXAikhsv6FjrPdcCZUFy$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118692229=
/7559972__;!!PvDODwlR4mBZyAb0!XpnKKmgtDbeWGvhHcYMtCjglwPRF-kjFEc9xwccJVSB_g=
qW6A_8gwdrAfS6U8pJgPlCMqgwYFXAikhsv6FjrPRMXOG6L$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Apr 06 2026 - 11:12:25 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST