Re: [AMMRL] trouble in helium recovery land...

From: Ulrike Werner-Zwanziger via groups.io <Ulli.Zwanziger=Dal.ca_at_groups.io>
Date: Mon, 6 Apr 2026 10:59:22 -0300

Hello Monika,

I second all the comments above, especially the one about the "Helium
tower" which we also call "Christmas tree". We have a lot of trouble
getting those to seal properly. A handheld helium detector is invaluable to
find those leaks, or any leak. We have a Restec Leak Detector, which has
become indispensable.

Good luck,

Ulli

On Mon, Apr 6, 2026 at 10:45 AM Mike Lumsden via groups.io wrote:

> Luke - in case you find it interesting, another approach to address the
> important question "Am I losing Helium" is to take a total helium inventory
> on, say, a weekly schedule. Putting this data into a spreadsheet and
> plotting over time can be quite informative!
>
> What I mean by taking an inventory is you sum all sources of liquid in
> your system and all sources of gas (in liquid equivalents). For an ideal
> system without any loss, the total inventory is constant until you purchase
> liquid. If you are only recovering gas, for off-site liquefaction, this is
> straight-forward with magnets the only source of liquid and gas in
> storage. Don't forget you also have gas in the magnets with gas volume
> ( total volume minus liquid volume). If you are also liquefying, then of
> course you must add in your production dewar and your mobile dewar(s) to
> the total inventory (with each dewar also being a source of gas). If you
> are recovering via a bag (as I do), my feeling is its best to manually pull
> the bag down before taking your inventory.
>
> Of course, as you point out, this all breaks down in the absence of
> accurate measurements of liquid volumes and those differential pressure
> meters seem to me to be quite coarse.
>
> I welcome all thoughts, concerns, ideas for improvement on this approach!
>
> Mike
>

--
Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way
PO BOX 15000
Halifax, NS B3H 4R2

e-mail: Ulli.Zwanziger_at_Dal.ca
Tel. 1 902 494 8085
FAX: 1 902 494 1310


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3352): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3352__;!!PvDODwlR4mBZyAb0!WS4zBOi8moDUb3Q2RXhMe2oaTmAuin=
aE871pE5iqCvS6bQyjvZ4uagu--37AKg7Ix9OQwePjTwcqGNaFGlwH3u-sFNJg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118639575=
/7559972__;!!PvDODwlR4mBZyAb0!WS4zBOi8moDUb3Q2RXhMe2oaTmAuinaE871pE5iqCvS6b=
QyjvZ4uagu--37AKg7Ix9OQwePjTwcqGNaFGlwH3tiuyBIv$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Apr 06 2026 - 07:01:10 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST