Re: [AMMRL] Noise complaints from Quantum Technology Helium Recovery System

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Fri, 3 Apr 2026 11:20:00 +0200

Hi,

your local DIY store should have some neoprene or similar foam wrapping which
is used for domestic heating tubes etc. - often this is available with a cut
along the length of the tube and sticky tape to close it again in order install
it in retrospective (no need to shut down and open the helium line)
This should help dampen the noise at least somewhat.

Cheers
Bert

> Am 02.04.2026 um 20:47 schrieb Fulton, Luke via groups.io:
>
> Hi Xu,
>
> I actually like the noise, those chirps reports on cold head operations,
> and their absence is an immediate alarm to investigate. I wouldn't recommend
> eliminating the noise entirely were it possible. Our dewar has a soft wrapping
> around it to dampen the volume, from touch/appearance I'd say it's all
> synthetics and foam. We sourced it directly from QT as part of the original
> install and would assume it's purchasable as a stand alone accessory. You
> could probably assemble something custom yourself to achieve similar results.
>
> I haven't noticed the chirping volume being any more severe than for helium
> cooled cryoprobes. Users might find it annoying from personal preference,
> but it doesn't rise to safety concern in my opinion. The HR3 compressor on
> the other hand is much louder. Thankfully it only turns on during high boil
> off scenarios which are either rare or predictable such as fills. I suggest/offer
> ear protection to users working long experiments while it's running.
>
> Kind regards,
> Luke
>
>
> Luke Fulton, PhD
> CHEM BLDG R003
> NMR Core Facility Director
> Unit 3060
> COR2E & Department of Chemistry
> 55 N Eagleville Road
> University of Connecticut
> Storrs, CT 06279
>
> email: fko24003_at_uconn.edu <mailto:fko24003_at_uconn.edu>
> (alias): luke.fulton_at_uconn.edu <mailto:luke.fulton_at_uconn.edu>
> mobile: (603) 953-5275
> Office: (860) 486-4069

> From: main_at_ammrl.groups.io on behalf of Xu, Xu via groups.io
> Sent: Tuesday, March 31, 2026 3:23 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Noise complaints from Quantum Technology Helium Recovery System
>
>
> Hi All,
> Anyone has complaints about the chirping noise from QT Helium Recovery System?
> How do you deal with this?
>
> Best,
> -Xu
>

Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3329): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3329__;!!PvDODwlR4mBZyAb0!ST_p08DvPXbU94jeu1EhHrUHk7O132=
d48K1IO912bbiBQrh-x9sTKfObyhkaEKH2Fd0UDtrRV7FC6bC-K8V6vg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118637955=
/7559972__;!!PvDODwlR4mBZyAb0!ST_p08DvPXbU94jeu1EhHrUHk7O132d48K1IO912bbiBQ=
rh-x9sTKfObyhkaEKH2Fd0UDtrRV7FC6bBzMaEu0A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Fri Apr 03 2026 - 07:36:00 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST