Re: [AMMRL] Probe Heater for VTU on Bruker EMXplus EPR

From: Jaya Nolt via groups.io <jaya=mrl.ucsb.edu_at_groups.io>
Date: Thu, 2 Apr 2026 15:54:36 -0700

Thanks Chuck.

We did take resistance measurements for the heater. For the safety sensor
1.8, and other pin measurements of 0.2 and 6.
I don't have a manual for the unit. I am hoping to get some more info from
Bruker soon on what the numbers should be.

best,
jaya

On Thu, Apr 2, 2026 at 11:23 AM Charles Hanson via groups.io wrote:

> Hi Jaya,
> I think the problem is either the heater is shorted and the VTU is drawing
> too much current or the VTU for some reason isn't able to provide current
> to the heater. So the unit applies a voltage and the current limiter kicks
> in because it is measuring too high a current and cuts the voltage back
> down to zero. In the manual it may specify what the heater
> resistance should be. Anyway I would measure this first.
>
> Cheers,
> Chuck
>
> On Thu, Apr 2, 2026 at 2:06 PM Jaya Nolt via groups.io wrote:
>
>> Hi Ammrl group,
>> I am looking for some troubleshooting help with our probe heater. Our EPR
>> is running Xenon with the VTU device.
>> Software says probe heater is OK however no heater power comes on, and it
>> stays at 0%. In the webpage I can turn on the heater briefly by clicking on
>> the options for 10%, 30% etc. I can see the heating effect and then the
>> percent drops back to 0.
>> One guess is that the heater safety sensor is not working properly.
>> The other guess is that the PID values have been comprimised.
>> In the software VTU controls there are no PID values listed in the table,
>> it is grayed out, and loading PID values is not supported. This seems odd
>> but I don't have any documentation otherwise. Also I cannot find anything
>> in the calibration files for VTU values.
>>
>> Thanks for any thoughts in advance.
>>
>> Jaya
>>
>>
>>
>>
>>
>
> --
> Charles Hanson
> Assistant Director, Magnetic Resonance Center
> Boston College
> Merkert 107
> 2609 Beacon Street
> Chestnut Hill MA 02467-3860
> Tel (617) 552-6903 (office)
>
>
> https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center.html
>
>
>
>

--
Jaya Nolt
R & D Engineer
Spectroscopy Facility
Materials Research Lab
Office: (805) 893-4997
Email: jaya_at_mrl.ucsb.edu
http://www.mrl.ucsb.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3332): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3332__;!!PvDODwlR4mBZyAb0!TK8ljbPSuc9y-Ok43aUC-4UU-u1JY5=
NgLWTIlGwS-J7liX0EojviB8t6u2kJD3qBBSNJt9figqDvphkvooM5KeDbMo8j$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118637968=
/7559972__;!!PvDODwlR4mBZyAb0!TK8ljbPSuc9y-Ok43aUC-4UU-u1JY5NgLWTIlGwS-J7li=
X0EojviB8t6u2kJD3qBBSNJt9figqDvphkvooM5KViv_K1h$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!TK8ljbPSuc9y-Ok43=



Received on Fri Apr 03 2026 - 07:36:54 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST