Dear TJ,

I ran into something similar when setting up a SOFAST experiment to run in
ICON. I got the '.../wave/0.0 does not exist' error. It occurred because
Topspin couldn't find the correct shaped pulse from the prosol table when
it ran 'getprosol'. To make it work, I had to add the correct shape,
power, and duration. I also had to modify the prosol relations file
('triple' in my case) to make sure the shaped pulse info (I think both
pulse and power) pointed to the correct line in the prosol table.

Good luck!

Best,
Peter

On Thu, Apr 2, 2026 at 2:23 PM Thusitha Jayasundera via groups.io wrote:

> Hello all--I have been trying to get a pseudo 2D experiment for a 31P
> kinetics experiment to work on automation (the 1D variant works great,
> thanks to the guidance of Eric Johnson from Bruker). After lots of help
> from several additional patient experts, I was able to make some
> progress--but run into an error when I include the ZGOPTIONS as instructed
> i.e. the line -DCALC_SP -DCALC_SPOFFS for ZGOPTIONS (to define the
> bandwidth and position of the selective pulse with CNST55 and CNST21)
> results in an error. The system is a 500 MHz Bruker NEO with a Prodigy
> probe...
>
> If I manually load the saved parameter set, I am able to run the
> experiment with no errors; the attached pulse sequence is displayed on
> Topspin. If I add the ZGOPTIONS, the two attached error messages appear
> (one in the acquisition window, the other when attempting to visualize the
> pulse sequence--which I also have attached here).
>
> I am assuming that I'm making a very simple and embarrassing mistake, and
> therefore would appreciate any available advice!
>
> With many thanks in advance,
>
> TJ.
>

--

*Peter Thompson, Ph.D.*
Lead Research Scientist, Macromolecular NMR, Molecular Education,
Technology and Research Innovation Center (METRIC)
Research Assistant Professor, Department of Molecular and Structural
Biochemistry
North Carolina State University
51 Polk Hall
Raleigh, NC 27695

Phone: 919.515.0836
Email: pmthomps_at_ncsu.edu
METRIC Website: https://research.ncsu.edu/metric/home
Consultation and Sample Submission: https://labs.ncsu.edu
Biochemistry Website: https://cals.ncsu.edu/molecular-and-structural-biochemistry/

“All electronic mail messages in connection with State business which are
sent to or received by this account are subject to the NC Public Records
Law and may be disclosed to third parties.”


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3339): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3339__;!!PvDODwlR4mBZyAb0!S4RKuwdMQurT8goqAXPMV8xsDK4s5M=
RnXTKPf2XRLwAcHHBQEzE9zyfBBqod5JO8X3dBnJPnOPgnNzqS_lUn8GwpLDBS$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118638239=
/7559972__;!!PvDODwlR4mBZyAb0!S4RKuwdMQurT8goqAXPMV8xsDK4s5MRnXTKPf2XRLwAcH=
HBQEzE9zyfBBqod5JO8X3dBnJPnOPgnNzqS_lUn8BejmoZU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Apr 03 2026 - 07:38:07 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST