Re: [AMMRL] Noise complaints from Quantum Technology Helium Recovery System

From: Matthias Brandl via groups.io <matthias.brandl=tum.de_at_groups.io>
Date: Thu, 2 Apr 2026 20:20:27 +0200

Dear Xu,

...earplugs? (sorry, could not resist...)

More seriously: The main source of noise in GM cryocoolers is flow noise
through the metal hoses between the compressor and the cold head. I
can't quite make out the manufacturer of the compressor from the picures
on QT's website, but a while ago, we were visited by a Sumitomo
representative. He said that they are now offering special "low noise"
hose packs. Probably quite expensive, but if noise is a serious issue
for you, maybe worth the cost.

Also, use the shortest hoses possible for your application and siting.
Shorter hoses are much quieter.

Best,
Matthias


On 31/03/2026 21:23, Xu, Xu via groups.io wrote:
> Hi All,
> Anyone has complaints about the chirping noise from QT Helium Recovery
> System?  How do you deal with this?
>
> Best,
> -Xu
>

--
Dr. rer. nat. Matthias Brandl (he/him or they/them)

Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)

Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen

Phone +49 89 289-52617

matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3323): https://ammrl.groups.io/g/main/message/3323
Mute This Topic: https://groups.io/mt/118637955/7559972
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/xyzzy [walt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-


Received on Thu Apr 02 2026 - 11:23:16 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST