Hi Thusitha,
NoMachine allows you to chose the connection protocol. While NX, NoMachine’s
own, is the default and works great, I personally don’t know how to limit it
to select users. However, if the server is linux, you can choose to connect via ssh
then the standard sshd_config settings in /etc/ssh/sshd_config or its equivalent
on the server side can be use to control users access. The relevant line in that
file is AllowUsers. Be sure to restart sshd after making changes.
John
> On Apr 24, 2025, at 2:27 PM, Thusitha Jayasundera via groups.io wrote:
>
> Hello all--I was wondering if any users who utilized NoMachine to remotely
> access the spectrometers could advise me on how they assured that only the
> nmrsu account was able to make the remote connection, blocking any other
> user names. This is likely a simple setting change, but unfortunately, I
> am unable to find it.
>
> Any guidance would be much appreciated!
>
John Decatur, Ph. D.
Senior Research Scientist, Director of Chemical Instrumentation
and Precision Biomolecular Characterization Facility
Department of Chemistry
3000 Broadway, Mailcode 3179
Columbia University
New York, NY 10027
212-854-2155 (voice)
212-932-1289 (fax)
jdd13_at_columbia.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2274): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2274__;!!PvDODwlR4mBZyAb0!TENapkz3TV9J8xF7kq6zrxCrCiw8Of=
BaSBQvhryBiIWQdcipELoucPkJjWwjGcXvwiIjubPzD4HP9zHhvzGhGYxcyKTC$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112438353=
/7559972__;!!PvDODwlR4mBZyAb0!TENapkz3TV9J8xF7kq6zrxCrCiw8OfBaSBQvhryBiIWQd=
cipELoucPkJjWwjGcXvwiIjubPzD4HP9zHhvzGhGfKyGKcY$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Apr 24 2025 - 11:46:25 MST