Re: [AMMRL] B-VT3200 and "overheating"-message

From: Nils Schlörer via groups.io <nils.schloerer=uni-jena.de_at_groups.io>
Date: Mon, 14 Apr 2025 10:11:25 +0000

Hello Bert,

yes. BCU was ruled out. Also, another probe (r.t.) was tested, too. With the same error.

Cheers,
Nils


________________________________________
> From: main_at_ammrl.groups.io on behalf of Spin-Doc
> Sent: Sunday, April 13, 2025 11:00 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] B-VT3200 and "overheating"-message

Hi Rainer,

any blockage of the VT air going to and through the probe has been ruled out?
Like a frozen, blocked BCU, snapped tube or the like?

Cheers
Bert

-> Sent via mobile <-

Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Web: http://www.spin-doc.net
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF

Am 13.04.2025 um 12:55 schrieb Platz Hirsch via groups.io:

Hi,

recently a friend had a nice challenge.
His equipment
- Avance III console
- BSMS/2
- B-VT 3200 as part of the BSMS/2

Since a few days he get's an overheating message. Switching cable and heater doesn't help,
neither switching the whole B-VT3200 rack.
Maybe the extra 48V power supply next to the B-VT 3200 at the rear side of the BSMS/2
might be the reason. An indication is an additional - non disturbing - error message about
a failing nitrogen evaporator.

Of course that friend could change the power supply on Monday. Nevertheless I am very curious.
Did you ever hear about the combination
failing 48V power supply - "overheating" message in the past?

Greetings

Rainer



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2223): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2223__;!!PvDODwlR4mBZyAb0!SVpLxRosUDULvivLxxMbukHcQABf_q=
8jKJOPsvZErXSO0L7SBrCs8OfZz9SAzW-BoaobUhL_kUkZotcCaRmkVP7G8YAp$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112239269=
/7559972__;!!PvDODwlR4mBZyAb0!SVpLxRosUDULvivLxxMbukHcQABf_q8jKJOPsvZErXSO0=
L7SBrCs8OfZz9SAzW-BoaobUhL_kUkZotcCaRmkVOzyQjQP$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Mon Apr 14 2025 - 05:34:10 MST

This archive was generated by hypermail 2.4.0 : Thu May 01 2025 - 13:19:57 MST