Re: [AMMRL] Issue with shims on Agilent DD2 400 MHz

From: Emily Crull via groups.io <emily.crull=mail.wvu.edu_at_groups.io>
Date: Wed, 02 Apr 2025 12:43:18 -0700

Hi everyone,

Good news update!!

After relieving all that back pressure, making a new 10% H2O in D2O standard,
and doing the 3D gradient shims, the Agilent 400 now has nice sharp peaks!

Now I'm working on getting those new shim maps and shims transferred to the
other users, but that's a software problem not an instrument problem.

Orders are being placed for more baffle and for the company to redo the plumbing
so we can fit the additional length.

Thanks so much to everyone who helped me fix the problem.

Emily
Director of Chemistry Shared Instrumentation
CRL 261  |  emily.crull_at_mail.wvu.edu
West Virginia University


-=-=-=-=-=-=-=-=-=-=-=-