Re: [AMMRL] INEPT

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 2 Apr 2025 11:44:54 +0200

Hi,

If all you need is a "1D out of a 2D" you can use the "proj" command: Flip
the HMBC you recorded to the N15 dimension being horizontal ("Rotate" button,
#12 icon from top), then type for example "proj(5)" to get a projection of
the 2D in experiment workspace #5. That may not have enough F2 resolution
enough to get coupling constants out though. You could enlarge resolution
by running a selHMBC instead of course.

Then again you can use the EXSIDE 2D (under "Sel2D") for just that (this amplifies
coupling constants to make them measurable again).

Alternatively, to get a "true 1D" 15N INEPT experiment you could try the following:
- select "Experiments - Other Nuclei - N15" from the menu
- type pslabel='NITROGEN' followed by "dg"
- select Tools - Study Clones - Edit Parlib, save as "NITROGE", choosing "Liquids"
Tabname and entering "Std 1D" as MenuLevel
- click "New Study"
- select "DEPT" experiment, doubleclick to edit parameters
- under "Default" select "No edit" for Multiplicity Editing and "Suppress" for Quaternary
- keep or remove shaped BB inversion pulses on "Pulsesequence" as you like
- enter REQexp='NITROGEN' customizepar('update') into the command line (sorry, DEPT
was not really meant to be customizable to other nuclei, so this function isn't under a button…)
- choose number of scans, spectral width etc., click Save
- Tools - Study Clones - Clone Current Study (save as DEPT-N15 for example)

Now you should have a button ready to use.

Best regards
Bert



> Am 31.03.2025 um 22:04 schrieb Vander Velde, David via groups.io :
>
> You should get their standard sequence if you just type inept on the command line in vnmrj.

> From: main_at_ammrl.groups.io on behalf of Shawn Wagner via groups.io
> Sent: Monday, March 31, 2025 12:08 PM
> To: main_at_ammrl.groups.io <mailto:main_at_ammrl.groups.io> <main_at_ammrl.groups.io <mailto:main_at_ammrl.groups.io>>
> Subject: [AMMRL] INEPT
>
> Does any have an INEPT sequence for Varian spectrometers? I haven't been
> able to find one posted anywhere. I am interested in obtaining 1D 15N spectra,
> but so far have been unsuccessful with the low SNR and long T1 values. =
> I am able to get a good 15N HMBC but cannot get a good 1D spectra even with reducing
> the acquisition window to the resonance frequency of the 15N nuclei in the molecule.
> My end goal is to try to measure 2J and 3J scalar couplings.
>
> -Shawn
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net <mailto:bert.heise_at_spin-doc.net>
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2195): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2195__;!!PvDODwlR4mBZyAb0!VAvdcOZErod3HmvrnPK4-swbKF1p0G=
nCMbWF2PyEzxxVMDIp7v127n5Zgn5wVRRKUcOoHOv71k7zt-zJaPqUVA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112011697=
/7559972__;!!PvDODwlR4mBZyAb0!VAvdcOZErod3HmvrnPK4-swbKF1p0GnCMbWF2PyEzxxVM=
DIp7v127n5Zgn5wVRRKUcOoHOv71k7zt-y7RgFC9g$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!VAvdcOZErod3Hmvrn=
PK4-swbKF1p0GnCMbWF2PyEzxxVMDIp7v127n5Zgn5wVRRKUcOoHOv71k7zt-xO2omWIQ$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Apr 02 2025 - 04:22:23 MST

This archive was generated by hypermail 2.4.0 : Thu May 01 2025 - 13:19:57 MST