[AMMRL] ICONNMR with ldap and sssd on CentOs7 #NMR_Aquisition_Software

From: Frank Holger Foersterling <holger_at_uwm.edu>
Date: Mon, 29 Apr 2024 14:33:23 -0700

AMMRLers,
I am using sssd krb5 and  LDAP in my lab to authenticate logins to our
Centos7 spectrometers and Fedora workstations.  I am keeping all user
information (group affiliation, numerical user and group id in an local LDAP
server. Most logins are authenticated by krb5 through our campus active
directory servers, but some NMR lab specific accountys also authenticate
through my LDAP server. With sssd that works for users logging into the
computers, and after installing the 32 bit sssd-client.i686 they can start
up Topspin.  However, ICONNMR (running under Topspin 3.7) sees only local
system users, but none of the users defined in LDAP . That is both in the
"Switch Users" dialog, and in the configuration window. All user profiles
are present in $TS/cont/instr/spec/inmrusers, but they are not shown.
Interestingly, the user currently logged into the system can unlock the
automation even if it is an LDAP user, but one can not switch to a different
non local user.

I looked at all the libraries Bruker recommends installing and made sure all
are installed, and if a 32 bit version was available, I also installed it but
to no avail.

ICONNMR will work fine and show all users if I use nslcd from the nss-pam-ldapd
package instead (with booth 32 and 64 bit versions installed), but I have not
been able to authenticate simultaneously  against our campus server and my
local directory server, which is why I tried sssd. My previous method was a
combination of krb5 and nis, which worked fine, but with the end of life for
Centos7 coming up and AlmaLinux no longer supportting nis I neeeeed to move
to an LDAP based solution.

It looks there was a similar question posted to this group a couple of years
ago by Charly Fry, but no solution was posted, but I hope someone can give me
a hint whether there is a solution to the problem. My suspicion is that there
is a 32 bit library missing, but I am not sure.

Thanks
Holger


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1205): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1205__;!!PvDODwlR4mBZyAb0!WYEHCdPYYMqB-9ZTEIk-YsNQWZhBCu=
dcNZhABo3pYuKwpcxIOPIl0XEMr6BbmrFkZqsXSRpWBq3GZdo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105810148=
/7559972__;!!PvDODwlR4mBZyAb0!WYEHCdPYYMqB-9ZTEIk-YsNQWZhBCudcNZhABo3pYuKwp=
cxIOPIl0XEMr6BbmrFkZqsXSRpWT2opHiI$
Mute #nmr_aquisition_software:https://urldefense.com/v3/__https://ammrl.gro=
ups.io/g/main/mutehashtag/nmr_aquisition_software__;!!PvDODwlR4mBZyAb0!WYEH=
CdPYYMqB-9ZTEIk-YsNQWZhBCudcNZhABo3pYuKwpcxIOPIl0XEMr6BbmrFkZqsXSRpWxQMHDwM=
$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Apr 29 2024 - 14:34:10 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:18 MST