Sometimes, but not always, you can see arcing by looking at the power reflected
from the probe on a scope. If you do see it there, then that is confirmation,
but if you don’t it doesn’t necessarily mean it isn’t arcing.
Tom Pratum, SOU
> On Apr 16, 2024, at 6:18 AM, Vernon M Russell via groups.io wrote:
>
> I’m getting sporadic spikes in the fid while running a cpmas sequence
> on a Neo (2.5 mm cpmas probe). I’ve tried easing up on the various
> power levels and pulse times but the sporadic nature makes this difficult to
> track down. Adjusting the spinning speed does seem to help. Does anyone have a
> general troubleshooting guide to see what is causing such spikes?
> Thanks,
> Vernon
>
> Vernon Russell
> vrussell_at_wlgore.com
> W. L. Gore & Associates, Inc.
> Together, improving life
> 501 Vieve's Way, P.O. Box 1320
> Elkton, MD 21922-1320
> USA
> gore.com
>
>
-=-=-=-=-=-=-=-=-=-=-=-