From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Fri, 12 Apr 2024 02:57:13 -0400
Dear Rainer,

First you said the magnet started drifting by 100 Hz/hour - I assume you mean here it drifted *down* in frequency first? And only sometime afterwards it started drifting upwards? Or is that its current speed drifting up? Indeed that may be its z0 coil discharging (a current built up in opposite direction to the main coil) faster than the main coil’s natural drift.

Anyway, it could be that the magnet is experiencing an overpressure in its helium vessel which should be vented as soon as possible: could you try to open one by one both helium turrets (maximum attention - magnet may be about to quench!!) and go in with e.g. an L-tube (the one used to precool a magnet with liquid nitrogen) to check for any air ice?

If the tube “sinks in” a bit slowly it may indicate air ice - pull back, warm up the tube with a heat gun and repeat. In case the magnet/helium vessel is/was under pressure, it may vent quite violently suddenly - be prepared for a sudden helium outrush (may look like a mini quench)…
If possible, have liquid helium at hand (you have a helium liquefier, right?) to refill quickly in case it loses a lot of helium.
After getting rid of the ice, go search for a (likely present) cracked O-ring or check vale (stuck open)…

Good luck!
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF


_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Online (#1133) | Reply To Group | Reply To Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Thu Apr 11 2024 - 23:57:37 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST