From: dennisonpr <dennison_at_uci.edu>
Date: Thu, 11 Apr 2024 10:24:59 -0700

Hi Rainer,

I had a drifting magnet which required periodic field adjustment. I used
expinstall to update parameter files, but found all 2H experiments
needed to b updated manually. This was an XwinNMR/DRX system.

In order to use expinstall I first copied all of the active parameter
files to the default install directory for the system, in this case
/u/exp/stan/nmr/par.avance

Regards,

Phil Dennison

On 20240411 07:39, Platz Hirsch wrote:
> Hi,
>
> although retired since about 5 years, there are nice challenges from time to time.
>
> The recent one within my former university gave me some questions.
>
> ==========
> Equipment
> ==========
> unshielded 360 MHz magnet, discharged to 300 MHz about 10 years ago
> helium level (dipstick measured): about 80%
>
> AV III console
>
> Windows workstation running TopSpin 3.2
>
> ==========
> Magnet behaviour
> ==========
>
> Since about one year, the helium level meter always shows the same value.
> To make the story short, with very high probability a lot of air
> condensed within the magnet,
> maybe due to a broken sealing.
> Now the magnet began to drift with about 100 Hz/hour.
> Very bad, presumably an ice bridge between one of the joints and a
> warmer part of the cryostate,
> but, of course, I am not sure.
>
> But now the unexpected things began ...
>
> ==========
> Drift direction
> ==========
>
> The magnetic field increases! It was possible to get the spectrometer
> back to operation
> by increasing the base frequency by 100 MHz (from 300.13 to 300.23)
> using cf.
>
> ==========
> TopShim
> ==========
>
> After changing the base frequency the data sets have to be corrected
> using expinstall.
> This works perfect for nearly all data sets stored within
>
>      .../exp/stan/nmr/par
>
> with the exception of topshim1d2H.
>
> This data set becomes *deleted*. Tried to manually install a copy from
> par.AVANCE, but running
> expinstall again to configure the parameters deletes this data set
> again. Only this one,
> nothing else.
> The data set for topshim1d1h is fine and produces the expected results.
> Didn't test the 3D data set.
>
> ==========
> user defined data sets
> ==========
>
> For ICON-NMR these are stored within
>      .../exp/stan/nmr/par/user
>
> expinstall doesn't touch these data sets. I tried a dirty trick and
> moved one data set from
>      .../par/user
>
> to
>
>      .../par
>
> but even in this case the user defined data set remains unchanged doing
> expinstall.
> I fear, the magnet is lost. Nevertheless I am very interested to
> understand this unexpected drift behaviour.
>
> Greetings
>
> Rainer
>
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1127): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1127__;!!PvDODwlR4mBZyAb0!QIaqahnGX_syJ5_VAhUa4KgnEZRuTs=
ImBFpeS-uczyB1H8galhPXC8gdQn7jQHlV1icncLlGtMak11hw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105467078=
/7559972__;!!PvDODwlR4mBZyAb0!QIaqahnGX_syJ5_VAhUa4KgnEZRuTsImBFpeS-uczyB1H=
8galhPXC8gdQn7jQHlV1icncLlGtIRDb9eb$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Thu Apr 11 2024 - 10:25:16 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST