(unknown charset) AMMRL: missing frequencies problem on VNMRS-600

From: (unknown charset) Jeff de Ropp <jsderopp_at_ucdavis.edu>
Date: Tue, 22 Apr 2008 11:40:19 -0700

Dear AMMRL,
we are having a problem with "missing frequencies" on the X-channel
of our VNMRS-600.

This shows up in a couple ways. First, we are seeing a large number
of U-shaped dips in the tune display that obscure our ability to
tune the probe for certain nuclei, for example Sn-119.
I have posted a screen capture of the problem:
http://www.nmr.ucdavis.edu/ammrl/mTuneScreen.pdf
We have tried the obvious things like switching probes and replacing
the cable to the probe. The problem is present with different probes.
The problem only occurs on the X-channel; proton is fine.

Second, if we use an external scope we can tune the probe fine to the
nuclei "obscured" in mtune, for example Sn119. However when we try
to acquire an Sn119 spectrum we see nothing, even for a concentrated
standard sample we normally see on one scan. So the problem is more
than just the tune display, certain frequency regions are now "gone"
and not observable.

Any ideas on solving this would be greatly appreciated.
Jeff